CVE-2025-0626
Jan. 31, 2025, 5:15 p.m.
7.5
High
Description
Contec Health CMS8000 Patient Monitor sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.
Product(s) Impacted
Product | Versions |
---|---|
UNKNOWN |
|
Weaknesses
CWE-912
Hidden Functionality
The product contains functionality that is not documented, not part of the specification, and not accessible through an interface or command sequence that is obvious to the product's users or administrators.
References
Tags
CVSS Score
CVSS Data
- Attack Vector: NETWORK
- Attack Complexity: HIGH
- Privileges Required: NONE
- Scope: UNCHANGED
- Confidentiality Impact: HIGH
- Integrity Impact: HIGH
- Availability Impact: HIGH
View Vector String
CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H
Date
- Published: Jan. 30, 2025, 7:15 p.m.
- Last Modified: Jan. 31, 2025, 5:15 p.m.
Status : Awaiting Analysis
CVE has been recently published to the CVE List and has been received by the NVD.
More infoSource
ics-cert@hq.dhs.gov
*Disclaimer: Some vulnerabilities do not have an associated CPE. To enhance the data, we use AI to infer CPEs based on CVE details. This is an automated process and might not always be accurate.