CVE-2026-0611

June 2, 2026, 8:16 p.m.

9.2
Critical

Description

Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.

Product(s) Impacted

Vendor Product Versions
Spacelabs Healthcare
  • Sentinel
  • 10.5.*, 11.*.*

Weaknesses

Common security weaknesses mapped to this vulnerability.

CWE-306
Missing Authentication for Critical Function
The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.

CVSS Score

9.2 / 10

CVSS Data - 4.0

  • Attack Vector: NETWORK
  • Attack Complexity: LOW
  • Attack Requirements: PRESENT
  • Privileges Required: NONE
  • User Interaction: NONE
  • Scope:
  • Confidentiality Impact: HIGH
  • Integrity Impact: HIGH
  • Availability Impact: HIGH
  • Exploit Maturity: NOT_DEFINED
  • CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

    View Vector String

Timeline

Published: June 2, 2026, 5:16 p.m.
Last Modified: June 2, 2026, 8:16 p.m.

Status : Deferred

CVE has been recently published to the CVE List and has been received by the NVD.

More info