CVE-2024-9832
Nov. 15, 2024, 1:58 p.m.
Tags
CVSS Score
Product(s) Impacted
UNKNOWN
Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Weaknesses
CWE-307
Improper Restriction of Excessive Authentication Attempts
The product does not implement sufficient measures to prevent multiple failed authentication attempts within a short time frame, making it more susceptible to brute force attacks.
CWE ID: 307Date
Published: Nov. 14, 2024, 9:15 p.m.
Last Modified: Nov. 15, 2024, 1:58 p.m.
Status : Awaiting Analysis
CVE has been recently published to the CVE List and has been received by the NVD.
More infoSource
productsecurity@baxter.com
CVSS Data
Attack Vector
LOCAL
Attack Complexity
LOW
Privileges Required
NONE
Scope
CHANGED
Confidentiality Impact
HIGH
Integrity Impact
HIGH
Availability Impact
HIGH
Base Score
Exploitability Score
Impact Score
Base Severity
CRITICALCVSS Vector String
The CVSS vector string provides an in-depth view of the vulnerability metrics.
View Vector StringCVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H