Description
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Product(s) Impacted
Product | Versions |
---|---|
UNKNOWN |
|
Weaknesses
Common security weaknesses mapped to this vulnerability.
CWE-1263
Improper Physical Access Control
The product is designed with access restricted to certain information, but it does not sufficiently protect against an unauthorized actor with physical access to these areas.
Tags
CVSS Score
CVSS Data - 3.1
- Attack Vector: LOCAL
- Attack Complexity: LOW
- Privileges Required: NONE
- Scope: CHANGED
- Confidentiality Impact: HIGH
- Integrity Impact: HIGH
- Availability Impact: HIGH
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H