Today > vulnerabilities   -   You can now download lists of IOCs here!

CVE-2024-48966

Nov. 15, 2024, 1:58 p.m.

CVSS Score

10.0 / 10

Product(s) Impacted

Unknown

Description

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Weaknesses

CWE-306
Missing Authentication for Critical Function

The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.

CWE ID: 306

Date

Published: Nov. 14, 2024, 10:15 p.m.

Last Modified: Nov. 15, 2024, 1:58 p.m.

Status : Awaiting Analysis

CVE has been recently published to the CVE List and has been received by the NVD.

More info

Source

productsecurity@baxter.com

CVSS Data

Attack Vector

NETWORK

Attack Complexity

LOW

Privileges Required

NONE

Scope

CHANGED

Confidentiality Impact

HIGH

Integrity Impact

HIGH

Availability Impact

HIGH

Base Score
10.0
Exploitability Score
3.9
Impact Score
6.0
Base Severity
CRITICAL
CVSS Vector String

The CVSS vector string provides an in-depth view of the vulnerability metrics.

View Vector String

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

References

https://www.cisa.gov/ productsecurity@baxter.com