CVE-2024-48966
Nov. 15, 2024, 1:58 p.m.
Tags
CVSS Score
Product(s) Impacted
Unknown
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Weaknesses
CWE-306
Missing Authentication for Critical Function
The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.
CWE ID: 306Date
Published: Nov. 14, 2024, 10:15 p.m.
Last Modified: Nov. 15, 2024, 1:58 p.m.
Status : Awaiting Analysis
CVE has been recently published to the CVE List and has been received by the NVD.
More infoSource
productsecurity@baxter.com
CVSS Data
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
Scope
CHANGED
Confidentiality Impact
HIGH
Integrity Impact
HIGH
Availability Impact
HIGH
Base Score
Exploitability Score
Impact Score
Base Severity
CRITICALCVSS Vector String
The CVSS vector string provides an in-depth view of the vulnerability metrics.
View Vector StringCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H