CVE-2024-48966
Nov. 15, 2024, 1:58 p.m.
10.0
Critical
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Product(s) Impacted
Product | Versions |
---|---|
Unknown |
|
Weaknesses
Common security weaknesses mapped to this vulnerability.
CWE-306
Missing Authentication for Critical Function
The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.
Tags
CVSS Score
CVSS Data - 3.1
- Attack Vector: NETWORK
- Attack Complexity: LOW
- Privileges Required: NONE
- Scope: CHANGED
- Confidentiality Impact: HIGH
- Integrity Impact: HIGH
- Availability Impact: HIGH
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Timeline
Published: Nov. 14, 2024, 10:15 p.m.
Last Modified: Nov. 15, 2024, 1:58 p.m.
Last Modified: Nov. 15, 2024, 1:58 p.m.
Status : Awaiting Analysis
CVE has been recently published to the CVE List and has been received by the NVD.
More infoSource
productsecurity@baxter.com
*Disclaimer: Some vulnerabilities do not have an associated CPE. To enhance the data, we use AI to infer CPEs based on CVE details. This is an automated process and might not always be accurate.