216.73.216.233

CVE-2019-25719

· Published 02/06/2026 14:16 · Modified 03/06/2026 16:16

Labels: CVE-2019-25719 2026-06-02CVE-2019-25719CWE-924[email protected]

Essential information

Published
02/06/2026 14:16
Modified
03/06/2026 16:16
Author
Creator
CVSS
8.8 HIGH (v3) 8.8 HIGH (v4.0)
CISA KEV
No
CWE
CVSS vector

CVSS metrics

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

NVD status

Status
Awaiting Analysis — CVE has been recently published to the CVE List and has been received by the NVD.
Source
[email protected]
NVD
View on NVD

Affected products (CPE)

ProductCPE
dräger / infinity acute care system cpe:2.3:a:dräger:infinity_acute_care_system:VG4.1.1-VG4.0.3:*:*:*:*:*:*:*
dräger / standalone infinity m540 cpe:2.3:a:dräger:standalone_infinity_m540:VG4.1.1-VG4.0.3:*:*:*:*:*:*:*

References